Important! Safety Notice and Recall of Certain Fluke Digital Clamp Meters
Dear Fluke Customer:
Fluke is voluntarily recalling certain 333, 334, 335, 336 and 337 Digital Clamp Meters that were manufactured between March 3, 2008 and February 27, 2009. The models and serial numbers affected are listed in the table below. If you own one of these clamp meters, please stop using it and send it back to Fluke for replacement. If you are not the primary user of the Fluke 33X Series Clamp Meter, please pass this notice along to the appropriate people within your organization
PLEASE NOTE: Models with the suffix R or the prefix S are NOT included in this recall.
Description of the problem
Through quality assurance tests, we have discovered a potential short circuit connection on the circuit board. This may lead to inaccurate voltage readings including a low or no voltage reading on a circuit energized with a hazardous voltage. This finding could create a hazardous situation if the user were to contact live voltage based on an erroneous instrument reading.
Actions
Within approx 10 working days you will receive a new unit from Fluke.
We are undertaking this voluntary recall because we place tremendous importance on our customers safety and the reputation of our products. Please accept our apologies for any inconvenience caused by this action.
|